113
Tel: 886.2.2175 2930 Email: [email protected]
Press
【
Tab
】
to switch to sliding block box, press
【
↑
】
,
【
↓
】
or rotate knob till pulse width or sidelobe level meets
the requirement.
Press
【
Tab
】
to switch to [Gaesarian Window β Para] to adjust setting of value β, till pulse width or sidelobe level
meets the requirement.
Press
【
Tab
】
to switch to [Pulse Span] to set response pulse width directly.
6) Set filtering function of time domain gate: press
【
Tab
】
to switch to [Window] option button, press
【
↑
】
,
【
↓
】
or rotate
knob to switch to [Gate Setting] option button, press
【
Tab
】
to switch to [Gating] check to, press
【
Click
】
to select the check
box and Stimulus filtering function of time domain gate, then set the function of time domain gate.
Press
【
Tab
】
to switch to [Start], [Stop] or [Center], [Span] to set gate border.
Press
【
Tab
】
to switch to [Gate Type] box, press
【
↑
】
,
【
↓
】
or rotate knob to select gate type: band pass, band
stop.
Press
【
Tab
】
to switch to [Gate Shape] box, press
【
↑
】
,
【
↓
】
or rotate knob to select gate form: minimum, standard,
wide, maximum.
7) If it is necessary to observe frequency domain response of the measured piece when the gate function is still effective,
press
【
Tab
】
to switch to [Gate Setting] option button, press
【
↑
】
,
【
↓
】
or rotate knob to switch to [Transform] option
button, press
【
Tab
】
to switch to [Transform], press
【
Click
】
to clear the check box and close time domain transformation
function.
8) Press
【
OK
】
in Input keypad to close Time Domain Transform dialog box after setting