An example of reflection Measure in band pass mode is provided below:
There is a pulse at 0 point on time axis, which represents the first discontinuous point (connector) met from the measuring port
output.
The subsequent pulse represents the second discontinuous point (connector) met by impulse signal.
Fig.8-15 Reflection Measure in Band Pass Mode
8.5.3.
Transmission Measure in Band Pass Mode
Horizontal axis data:
It represents transmission delay when passing the measured piece.
Vertical axis data:
In the format of logarithm amplitude, it represents transmission loss or transmission gain.
In the format of linear amplitude, it represents transmission coefficient.
An example of surface filter (SAW) transmission Measure in band pass mode is provided below:
Fig.8-16 Transmission Measure in Band Pass Mode
Three transmissions in surface acoustic wave (SAW) filter can be found from Measure results: the signal is reflected back to
input port by output port and then reflected back to output port again.
Pulse closest to 0 time point represents transmission time of the shortest path between input and output of the measured
piece; however, it is neither the maximum pulse definitely nor representing the expected transferring path.