Ripple Limit (on/OFF)
To enable the ripple test function if checked, to disable the ripple test function if unchecked.
Ripple Line (on/OFF)
To display the fluctuation line of the activated trace if checked, to close the display if unchecked,but it does not influence the ripple
test function.
Sound ON Fail (on/OFF)
If checked, the buzzer will send out alarm sound when the data point test on the trace fails.
Fail Sign (on/OFF)
If checked, the FAIL mark will be displayed on the screen when the data point test on the trace fails.
Show Table
Click it to display and edit the ripple limit table. It will be shifted to disabled state after the table is opened.
Hide Table
Click it to close the display of ripple limit table.It will be shifted to disabled state after the table is closed.
Export
Click it, storage dialog box will appear. Users could select storage route and save the file with *.csv format.
Import
Click it, storage dialog box will appear. Select the storage route of the file to be opened. It can open the file of *.csv format.
Val Type
Click the drop-down box to select the ripple value type: none
、
absolute
、
margin.The default setup of the analyzer is absolute value.
Ripple Band
Click the drop-down box to select the number of fluctuation limit segments to be opened and it can open 12 segments at best.
6.6.3.
Display ripple test result
If the result of ripple test passes, the word “PASS” will appear on the upper left side of the screen, otherwise, the word “ FAIL” will
appear on the screen. If multiple trace tests are made at the same time, the word “ PASS” or “ FAIL” for each trace will appear
respectively at the same time. The stimulus ranges of different ripple limit segments can overlap with different fluctuation limits.