Reference Manual
00809-0100-4738, Rev DA
May 2006
Rosemount 3095FB
5-9
1.
Inspect all cover and housing (non-process-wetted) O-rings and
replace if necessary. Lightly grease with silicone lubricant to ensure a
good seal.
2.
Carefully tuck the cable connector completely inside the internal
shroud. To do this, turn the shroud and cable counterclockwise one
rotation to tighten the cable.
Figure 5-5. Transmitter
Reassembly
3.
Lower the electronics housing onto the module. Guide the internal
shroud and cable through the housing and into the external shroud.
4.
Turn the housing clockwise to fasten it to the module.
5.
Thread the housing completely onto the sensor module. The housing
must be no more than one full turn from flush with the sensor module
to comply with explosion proof requirements.
6.
Tighten the housing rotation set screw using a hex wrench.
3095\016ab,031ab
To prevent damage to the cable connector, watch the cable
and shroud as you attach the housing to the module. Make
sure the cable connector does not slip out of the internal
shroud and begin to rotate with the housing. Reinsert the
cable connector into the shroud if it escapes before the
housing is fully fastened.