Reference Manual
00809-0100-4738, Rev DA
May 2006
Rosemount 3095FB
5-6
Remove the transmitter
from Service
Once a transmitter is determined to be inoperable, remove it from service.
Be aware of the following:
•
Isolate and vent the process from the transmitter before removing the
transmitter from service.
•
Remove all electrical leads and conduit.
•
Detach the process flange by removing the four flange bolts and the
two alignment screws that secure it.
•
Do not scratch, puncture, or depress the isolating diaphragms.
•
Clean isolating diaphragms with a soft rag and a mild cleaning solution,
and rinse with clear water.
•
Whenever you remove the process flange or flange adapters, visually
inspect the Teflon O-rings. Replace the O-rings if they show any signs
of damage, such as nicks or cuts. If they are undamaged, you may
reuse them.
Remove the Terminal
Block
Electrical connections are located on the terminal block in the compartment
labeled “FIELD TERMINALS.” Remove the housing cover (see Figure 5-1).
Loosen the two small screws located at the 9 o'clock and 4 o'clock positions,
and pull the entire terminal block out to remove it.
Figure 5-1. Removing the
Terminal Block Housing Cover
Remove the Electronics
Board
The transmitter electronics board is located in the compartment opposite the
terminal side. To remove the electronics board, perform the following
procedure:
1.
Remove the housing cover opposite the field terminal side.
2.
Loosen the two captive screws that anchor the board to the housing.
3.
The electronics board is electrostatically sensitive; observe handling
precautions for static-sensitive components.