Mount the Transmitter
Figure 2-2 illustrates a typical 3095 installation site. Major components of the
installation are identified in these figures.
Figure 2-2. Typical Rosemount 3095FB Installation Site
The 3095FB transmitter total weight varies depending on the components
ordered (see Table 2-1). This weight must be securely supported.
Table 2-1. Transmitter Weight
Mounting Brackets
Optional mounting brackets available with the 3095FB facilitate mounting to a
panel, wall, or 2-inch pipe. The bracket option for use with the Coplanar flange
is 316 SST with 316 SST bolts. Figure 2-3 shows bracket dimensions and
mounting configurations for this option. When installing the transmitter to one
of the mounting brackets, torque the bolts to 125 in-lb.
3095FB
RTD Connector
Process
Connections
RTD Cable
RTD Assembly
Conduit Opening
for Connections
to RTU and
Power Supply
3095-3095f22a
Flow
Component
Weight
3095FB Without Options
6 lb (2,7 kg)
LCD Meter for Aluminum Housing
0.5 lb (0,2 kg)
SST Mounting Bracket for Coplanar Flange
1.0 lb (0,5 kg)
12 ft (3,66 m) cable
0.5 lb (0,2 kg)
24 ft (7,32 m) cable
2.2 lb (1,0 kg)