![Rosemount 3095FB Reference Manual Download Page 116](http://html.mh-extra.com/html/rosemount/3095fb/3095fb_reference-manual_1487388116.webp)
Reference Manual
00809-0100-4738, Rev DA
May 2006
Rosemount 3095FB
4-6
USER INTERFACE
SOFTWARE SCREENS
This section goes over each of the screens found in the 3095FB User
Interface Software.
File Menu
New Config
Create a completely new configuration file from scratch.
Open Config…
Load a configuration file that has previously been created and saved on the
computer.
Save Config
Save the current configuration file under the same file name.
Save Config As…
Save the current configuration but with a new file name.
Exit
Shut down and exit the 3095FB UI Software.
Setup Menu
Setup > Host Communications…
Port Parameters:
•
Port:
Specifies the serial port on the PC that the RS232-485 converter
connects to. COM1 is the most common port in most laptop PCs.
•
Bit Rate:
Specifies the baud rate for the PC serial communication port.
This baud rate must match the transmitter’s baud rate as selected by
the jumper positions on the electronics comm. board (see page 2-12).
•
CTS Delay:
Specifies the delay before a Clear-To-Send message is
sent out. (Valid range is 0 – 9999 ms, recommended value is 25 ms.)
•
Hardware CTS Enabled:
Enable / Disable hardware Clear-To-Send.
Poll Parameters:
•
Retries:
Specifies the number of retries a Modbus command is resent
before the computer declares failed communication. (Valid range is 0
-99, recommended value is 3.)
•
Reply Timeout:
Specifies the maximum response duration before the
computer declares failed communication. (Valid range is 0 – 9999 ms,
recommended value is 1000 ms.)
•
Interpoll Delay:
(Valid range is 0 – 99 ms, recommended value is 10
ms.)
•
Poll Interval:
(Valid range is 0 – 999,999 ms, recommended value is
2000 ms.)
•
Extra Cmd Time:
(Valid range is 0 – 9999 ms, recommended value is
500 ms.)