Accura 2300/2350 User Guide
Appendix C Accuracy/Reliability
ⓒ
2011 Rootech Inc. All Rights Reserved
Page 215
IEC62053-22, Electricity Meter Equipment: active energy (Class 0.5S and Class 0.2S)
Current Range
Power Factor
Class 0.5S
Class 0.2S
0.01 I
n
- 0.05 I
n
1.0
±
1.0 %
±
0.4 %
0.05 I
n
- I
max
1
1.0
±
0.5 %
±
0.2 %
0.02 I
n
- 0.1 I
n
0.5 inductive
±
1.0 %
±
0.5 %
0.8 capacitive
±
1.0 %
±
0.5 %
0.1 I
n
- I
max
0.5 inductive
±
0.6 %
±
0.3 %
0.8 capacitive
±
0.6 %
±
0.3 %
0.25 inductive
±
1.0 %
±
0.5 %
0.5 capacitive
±
1.0 %
±
0.5 %
1. Maximum current I
max
= 1.2 I
n
, where rated current I
n
= 5A, 30A, 60A, 100A, 125A, 250A, 400A, 630A, 800A.
Reliability
IEC61326, Electrical Equipment for Measurement, Control, and Laboratory- EMC
Standard
Compliance
Accura 2300/2350
IEC61000-4-2
Electrostatic discharge[ESD] immunity
4kV/8kV contact/air
12kV/20kV
IEC61000-4-3
Radiated, radio-frequency, electromagnetic
field immunity
10V/m, 80M - 2700MHz
10V/m, 80M - 2700MHz
IEC61000-4-4
Electrical fast transient/burst immunity
2kV
4kV
IEC61000-4-5
Surge immunity
1kV/2kV,
line-to-line/line to earth
2kV/4kV
IEC61000-4-6
Immunity to conducted disturbances, induced
by radio-frequency fields
10V
10V
IEC61000-4-8
Power frequency magnetic field immunity
30A/m
30A/m
IEC61000-4-11 Voltage dips, short interruptions and voltage
variations immunity
0.5 cycle,
each polarity 100%
0.5 cycle,
each polarity 100%
Summary of Contents for Accura 2300 Front
Page 221: ......