Operating the Instrument
R&S
®
FSMR3000
78
Getting Started 1179.0100.02 ─ 01
Icons for individual settings
A bullet next to the setting indicates that user-defined settings are used, not auto-
matic settings. A green bullet indicates this setting is valid and the measurement
is correct. A red bullet indicates an invalid setting that does not provide useful
results.
Common settings
The channel bar above the diagram not only displays the channel-specific set-
tings. It also displays information on instrument settings that affect the measure-
ment results even though it is not immediately apparent from the display of the
measured values. This information is displayed in gray font and only when appli-
cable for the current measurement, as opposed to the channel-specific settings
that are always displayed.
The following types of information are displayed, if applicable.
Table 7-2: Common settings displayed in the channel bar
"SGL"
The sweep is set to single sweep mode.
"Sweep
Count"
The current signal count for measurement tasks that involve a specific number
of subsequent sweeps
(see "Sweep Count" setting in "Sweep settings" in the User Manual)
"TRG"
Trigger source
(for details see "Trigger settings" in the User Manual)
●
EXT
: External
●
IFP
: IF power (+trigger bandwidth)
●
PSE
: Power sensor
●
RFP
: RF power
●
SQL
: Squelch
●
TIM
: Time
●
VID
: Video
"6dB"/"RRC"
/"CHN"
Filter type for sweep bandwidth
(see "Bandwidth settings" in the User Manual)
"PA"
/
Ext
"PA"
The preamplifier is activated. / Data compensation is performed using data
from the (optional) external preamplifier.
Understanding the Display Information