Trying Out the Instrument
R&S
®
FSMR3000
51
Getting Started 1179.0100.02 ─ 01
5. Tap the "Calibration Frequency RF" option. Leave the frequency at the default
64
MHz, with a narrowband spectrum.
The calibration signal is now sent to the RF input of the R&S
FSMR3. By
default, a continuous frequency sweep is performed, so that the spectrum of
the calibration signal is now displayed in the standard level versus frequency
diagram.
Figure 6-1: Calibration signal as RF input
Instrument warmup time
Note that the instrument requires an initial warmup time after switching it on.
A message in the status bar ("Instrument warming up...") indicates that the
operating temperature has not yet been reached. Wait until this message is
no longer displayed before you start a measurement.
To optimize the display
To optimize the display for the calibration signal, we will adjust the main measure-
ment settings.
1. Set the center frequency to the calibration frequency:
a) Tap the "Overview" softkey to display the configuration "Overview".
Measuring a Basic Signal in the Spectrum Application