Trying Out the Instrument
R&S
®
FSMR3000
64
Getting Started 1179.0100.02 ─ 01
2. Tap the diagram near the first peak and drag your finger to the opposite corner
of the zoom area. A white rectangle is displayed from the point where you tap-
ped to the current position.
Figure 6-12: Defining the zoom area
When you remove your finger, the zoom area is enlarged in a second
(sub-)window.
Figure 6-13: Zoomed display around a peak
3. In
, the enlarged peak is represented by a very thick trace. This is
due to the insufficient number of sweep points. The missing sweep points for
the zoomed display are interpolated, which provides poor results. To optimize
the results, we will increase the number of sweep points from the default 1001
to 32001.
a) Press the [Sweep] key on the front panel.
b) Tap the "Sweep Config" softkey in the "Sweep" menu.
c) In the "Sweep Points" field, enter
32001
.
d) Press the RUN SINGLE key on the front panel to perform a new sweep
with the increased number of sweep points.
Zooming into the Display