ENV216
Operating Manual
3560.6550.12 -01-
4.5
E-1
3 Reference ground rail
ground rail
The ground rail that is included on the side of the
device is used exclusively for connecting a reference
ground. A broad metal foil can be placed over it to
create a low-inductance connection with the test
setup.
4 Grounding conductor connection
Grounding
conductor
connection
The connection, a 6-mm-threaded bolt with butterfly
nut is used only for protective grounding. It is not
suitable for use as a reference ground.
Caution!
The instructions in chapter 1 "Safety directions" and
chapter 3 "Preparation" must be followed at all times
5 EXTERNAL POWER SUPPLY
External Power Supply Connection for the plug-in power supply. The control
circuitry of the ENV216 is powered via the
“EXTERNAL POWER SUPPLY" connection, if the
ENV216 is to be used to conduct measurements of
a.c. voltages below 100 VAC or on d.c. voltages up
to 50 VDC. (for more information see also chapter 3,
section entitled "Operating with non-standardized
voltages")
NUR AN NETZE MIT
SCHUTZLEITER ANSCHLIESSEN
CONNECT TO MAINS SUPPLY
WITH PROTECTIVE EARTH !
EXT. POWER
SUPPLY
10 ... 18 VDC
250 mA
47 ... 63 Hz
I = 16 A
V = 50 V
DC. max.
V = 250 V
eff. max.
max.
REMOTE CONTROL
Ca u t i on
For laboratory use by
qualified personnel
CAUTION
ACHTUNG
(High leakage current of approx. 800 mA)
(Hoher Ableitstrom -ca. 800 mA)
Put unit into operation in the following order:
1. connect artificial mains network to protective earth
using additional connector
2. Insert power connector in socket of supplying network
Put unit out of operation in the following order:
1. Disconnect artificial mains network from supplying network
2. Disconnect additional protective earthing
Inbetriebnahme in folgender Reihenfolge:
1. Netznachbildung mit zusätzlichem Schutzleiter verbinden
2. Netzstecker mit dem zu speisenden Netz verbinden
Außerbetriebnahme:
1. Netznachbildung vom speisenden Netz trennen
2. zusätzliche Schutzleiterverbindung trennen
1
2
4
5
3
Summary of Contents for 3560.6550.06
Page 2: ...Operating Manual ENV216 3560 6550 12 01 1 2 E 1 ...
Page 4: ...Operating Manual ENV216 3560 6550 12 01 1 4 E 1 ...
Page 14: ......
Page 16: ...Operating Manual ENV216 3560 6550 12 01 1 6 E 1 ...
Page 18: ......
Page 20: ...Operating Manual ENV216 3560 6550 12 01 2 2 E 1 Block diagram ...
Page 30: ......
Page 36: ......