Operating Manual
ENV216
3560.6550.12 -01-
3.4
E-1
Operation with non-standardized voltages
For measurement tasks on voltages that are outside the normal mains voltage ranges (e.g. <100 VAC)
and for operation on d.c. (max. 50 VDC) the ENV216 control circuit is powered externally by a plug-in
power supply.
Caution!
The d.c. voltage must not exceed 50 VDC!
The plug-in power supply is connected to the "EXT. POWER SUPPLY" socket (13). The input is reverse
voltage protected.
EXT POWER SUPPLY socket
Caution!
When external power supplies are used to power the device with the auxiliary DC
voltage (SELV) the requirements for reinforced/double insulation in accordance with
DIN/EN/IEC 61010 (UL 3111,CSA C22.2 No. 1010.1) or DIN/EN/IEC 60950
(UL 1950, CSA C22.2 No. 950) must be satisfied.
NUR AN NETZE MIT
SCHUTZLEITER ANSCHLIESSEN
CONNECT TO MAINS SUPPLY
WITH PROTECTIVE EARTH !
47 ... 63 Hz
I = 16 A
V = 50 V
DC. max.
V = 250 V
eff. max.
max.
Ca ut i on
For laboratory use by
qualified personnel
ACHTUNG
(Hoher Ableitstrom -ca. 800 mA)
Inbetriebnahme in folgender Reihenfolge:
1. Netznachbildung mit zusätzlichem Schutzleiter verbinden
2. Netzstecker mit dem zu speisenden Netz verbinden
Außerbetriebnahme:
1. Netznachbildung vom speisenden Netz trennen
2. zusätzliche Schutzleiterverbindung trennen
1
Summary of Contents for 3560.6550.06
Page 2: ...Operating Manual ENV216 3560 6550 12 01 1 2 E 1 ...
Page 4: ...Operating Manual ENV216 3560 6550 12 01 1 4 E 1 ...
Page 14: ......
Page 16: ...Operating Manual ENV216 3560 6550 12 01 1 6 E 1 ...
Page 18: ......
Page 20: ...Operating Manual ENV216 3560 6550 12 01 2 2 E 1 Block diagram ...
Page 30: ......
Page 36: ......