Measurement setup
The ENV216 Two-Line V-Network satisfies the requirements of interference measurement regulations
VDE 0876, CISPR 16-1, and the U.S. FCC (Rules and Regulations Part 15J).
The measurement setups prescribed in the individual standards are largely similar. VDE regulation
0877 provides a detailed description.
Devices under test that are not equipped with permanently attached connecting wires are connected to
the network using a 1 meter long, unshielded wire. For devices under test equipped with a permanently
attached cable that is longer than 1 meter, the cable is laid out in a folded configuration.
Item
Description
1
Metal wall, at least 2 m x 2 m
2
Equipment under test
3
Cable in folded configuration
4
Two-Line V-Network
5
Shielded connection cable
6
Test receiver
B
Reference ground connection
M
Test receiver connector
P
Test device connector