ENV216
Operating Manual
3560.6550.12 -01-
3.9
E-1
Sample measurement
The figure shows the measurement setup for measuring the RFI voltage of a handheld drill. In this ex-
ample, the handheld drill is connected to three metal foils.
Item
Description
1
Metal wall, at least 2 m x 2 m
2
Equipment under test
3
Connection cable
4
Connection cable running separately to the hand simulation
5
Two-Line V-Network
6
Metallic collar capable being gripped
7
Test receiver
A
Power supply connection
B
Reference ground connection, low-inductance (e.g. brass plate,
0.2 m)
P
Test device connector
H
Artificial Hand connector
M
Test receiver connector
D
Metal foils