ENV216
Operating Manual
3560.6550.12 -01-
2.1
E-1
2 Brief introduction
Introduction
The ENV216 compact Two Line V-Network is used to measure noise voltage on mains-dependent con-
sumers.
Its principal tasks are:
•
To supply the equipment under test with mains voltage
•
To provide a standardized load impedance
•
Defined transmission of the noise voltage generated by the equipment under test to the EMI test
receiver
•
Isolation of the test circuit from interference of the power source.
The ENV216 Two Line V Network is constructed using air-core inductances (50 µH and 250 µH) and
conforms with the recommendations of VDE0876 and CISPR 16-1.
Other features of the equipment include:
•
Artificial Hand
•
10-dB-attenuator in the RF path
•
Switchable 150-kHz-highpass filter
•
Pulse voltage limiter at the measurement output
•
Remote control port
•
External power supply via plugged power pack for tests on non-standardized a.c and d.c. voltages
The network may be operated manually or by remote control with TTL levels.