ENV216
Operating Manual
3560.6550.12 -01-
3.3
E-1
Mains connection
Connection to the mains power supply is provided by a permanently attached mains cable.
The device is not equipped with an ON / OFF switch.
Caution!
When beginning operation, the sequence below must be followed strictly:
Connect network to auxiliary protective grounding
Connect mains plug to the mains power supply
Switching off:
Unplug network from the mains power supply
Disconnect auxiliary grounding connection
Caution!
Due to the high level of earth leakage current, the device cannot be operated on
power supplies protected by residual current devices (RCDs).
The power source must be protected by a circuit breaker with a maximum rating of
16 A characteristic B or C.
If the phase connection is made correctly, LED L1 (1) must light up.
If it does not:
Ø
For mains plugs with symmetrical grounding connection (e.g. Schuko plugs) the plug's polarity must
be reversed.
Ø
For mains plugs with asymmetrical grounding connection, phase L1 and neutral N must be swapped
by reconnecting.
L1 LED, phase indicator
The plug on the mains cable can be adapted using a country-specific adapter. However, it must be
ensured that the adapter has a power rating of at least 16 A. Assembly must be carried out exclusively
by qualified technicians.
1
Summary of Contents for 3560.6550.06
Page 2: ...Operating Manual ENV216 3560 6550 12 01 1 2 E 1 ...
Page 4: ...Operating Manual ENV216 3560 6550 12 01 1 4 E 1 ...
Page 14: ......
Page 16: ...Operating Manual ENV216 3560 6550 12 01 1 6 E 1 ...
Page 18: ......
Page 20: ...Operating Manual ENV216 3560 6550 12 01 2 2 E 1 Block diagram ...
Page 30: ......
Page 36: ......