![Rohde & Schwarz 3560.6550.06 Operating Manual Download Page 33](http://html.mh-extra.com/html/rohde-and-schwarz/3560-6550-06/3560-6550-06_operating-manual_1477795033.webp)
ENV216
Operating Manual
3560.6550.12 -01-
4.3
E-1
3 RF OUTPUT
RF OUTPUT
The "RF OUTPUT" socket is the RFI voltage meas-
urement output for connecting the test receiver. The
RFI voltage spectrum of the selected path is attenu-
ated at the "RF OUTPUT" by 10 dB. The 10-dB-
attenuator and a transient suppressor are perma-
nently inserted. A 150-kHz-highpass filter can addi-
tionally be switched into the path either manually
or by remote control. The socket is an "N" type.
The output impedance is 50
Ω
.
4 ARTIFICIAL HAND
ARTIFICIAL HAND
The "ARTIFICIAL HAND" enables the effects of the
human hand to be simulated. Metal housing parts of
the equipment under test that are usually touched
with the hand can be connected to this jack. The
hand simulation consists of connection in series of a
510-
Ω
-resistor and a 220-pF-capacitor. The jack is a
4-mm-terminal clamp.
(For more information, see also "Sample measure-
ment" in chapter 3)
TWO-LINE V-NETWORK
ENV216
3560.6550.06
1
2
3
4
Summary of Contents for 3560.6550.06
Page 2: ...Operating Manual ENV216 3560 6550 12 01 1 2 E 1 ...
Page 4: ...Operating Manual ENV216 3560 6550 12 01 1 4 E 1 ...
Page 14: ......
Page 16: ...Operating Manual ENV216 3560 6550 12 01 1 6 E 1 ...
Page 18: ......
Page 20: ...Operating Manual ENV216 3560 6550 12 01 2 2 E 1 Block diagram ...
Page 30: ......
Page 36: ......