MultiView tab
An additional tab labelled "MultiView" provides an overview of all active channels
at a glance. In the "MultiView" tab, each individual window contains its own chan-
nel bar with an additional button. Tap this button to switch to the corresponding
channel display quickly.
Icons in the channel bar
The star icon ( ) on the tab label indicates that the displayed trace no longer
matches the current instrument settings. This may be the case, for example, if a
trace is frozen and the instrument settings are changed. As soon as a new mea-
surement is performed, the icon disappears.
The exclamation mark ("!" or ) icon indicates that an error or warning is available
for that measurement channel. This is particularly useful if the MultiView tab is
displayed.
The icon indicates the currently active channel during an automatic measure-
ment sequence (
Sequencer
functionality).
Beneath the channel name, information on channel-specific settings for the mea-
surement are displayed in the
channel bar
. Channel information varies depend-
ing on the active application.
In addition to the channel-specific settings, the channel bar above the diagram
also displays information on instrument settings that affect the measurement
results even though this is not immediately apparent from the display of the mea-
sured values. This information is displayed only when applicable for the current
measurement.
The following table contains all labels that might be displayed in the Spectrum
and Receiver applications.