Trying Out the Instrument
R&S
®
FSWP
68
Getting Started 1177.5756.02 ─ 13
5. View the .dat file that has been created with your preferred program.
5.4
Measurement Configuration
The Phase Noise application provides several tools that make the measurement
process easier to handle or allow you to customize the measurement configura-
tion according to your needs.
.................................................................................68
The Half Decade Configuration Table
............................................................. 69
Pulsed Measurement Configuration
................................................................70
........................................................................................... 71
..........................................................................................72
5.4.1
Frequency Configuration
How to define the measurement range
The measurement range defines the carrier frequency offset over which the
phase noise measurement takes place.
1. Enter the "Overview" dialog box, and from there, enter the "Noise Config" dia-
log box.
2. Select the "Noise Config" tab.
3. Define the measurement range with the "Start Offset" and "Stop Offset" prop-
erties.
The application adjusts the measurement range. You can see this change, for
example, in the noise diagram result display.
Note
: In case of baseband measurements, the measurement range is defined
by an absolute start and stop frequency instead of offsets.
For more information on the measurement range, refer to the R&S
FSWP
user manual.
Measurement Configuration