![Rohde & Schwarz 1322.8003K08 Getting Started Download Page 31](http://html1.mh-extra.com/html/rohde-and-schwarz/1322-8003k08/1322-8003k08_getting-started_1477775031.webp)
Preparing for Use
R&S
®
FSWP
31
Getting Started 1177.5756.02 ─ 13
To suppress electromagnetic radiation during operation:
●
Use high-quality shielded cables, for example, double-shielded RF and LAN
cables.
●
Always terminate open cable ends.
●
Ensure that connected external devices comply with EMC regulations.
Signal input and output levels
Information on signal levels is provided in the data sheet. Keep the signal levels
within the specified ranges to avoid damage to the R&S
FSWP and connected
devices.
Considerations for Test Setup