Chapter 4 Common Modulation Output
RIGOL
User’s Guide for DG5000
4-11
Amplitude Shift Keying (ASK)
When using ASK (Amplitude Shift Keying) modulation, you can configure the
generator to make its output amplitude “shift” between two preset amplitude values
(“carrier amplitude” and “modulating amplitude”). The “shift” rate (ASK Rate) is
determined by the signal level inside the instrument or from the [ExtTrig] connector at
the rear panel.
To Select ASK Modulation
Press Mod
Type
ASK to enable ASK function.
When Mod is enabled, Sweep or Burst will be disabled automatically (if
enabled currently).
After ASK is enabled, the instrument will generate ASK waveform with the
currently specified carrier and modulating waveform.
To Select Carrier Waveform Shape
ASK carrier waveform shape could be Sine, Square, Ramp or arbitrary waveform. The
default is Sine.
Press the front panel key Sine, Square, Ramp or (Arb
Select Wform) to
select desired carrier waveform shape.
Pulse (basic waveform), Noise and DC could not be used as carrier waveform.
To Set Carrier Amplitude
After selecting carrier waveform shape, press Ampl/HiLevel to highlight “Ampl”, and
then use the numeric keyboard or knob to input desired amplitude. The available
amplitude range is limited by Resistant and Freq/Period. Please refer to the
“Output Characteristic” described in “
Specifications”.
www.
.com
1.800.561.8187
Summary of Contents for DG5071
Page 2: ...www com information itm com 1 800 561 8187...
Page 46: ...www com information itm com 1 800 561 8187...
Page 78: ...www com information itm com 1 800 561 8187...
Page 114: ...www com information itm com 1 800 561 8187...
Page 126: ...www com information itm com 1 800 561 8187...
Page 134: ...www com information itm com 1 800 561 8187...
Page 200: ...www com information itm com 1 800 561 8187...
Page 210: ...www com information itm com 1 800 561 8187...