Chapter 8 FH Output (Option)
RIGOL
User’s Guide for DG5000
8-9
FH Sequence
As shown in the figure above, if there are 8 frequencies available in a FH system
expressed as numbers from 1 to 8, then {8,1,2,4,6,3,5,7,6,1,2} is a simple FH
sequence. The sequence determines that the output sequence of the frequency is
{f
8
,f
1
,f
2
,f
4
,f
6
,f
3
,f
5
,f
7
,f
6
,f
1
,f
2
}.
Figure 8-4 FH Sequence in FH Map
1.
Built-In Sequence
Press Load Sequence
BuiltIn to select the three built-in sequences.
Normal: a normal increasing sequence with the length as n (n is equal to the
number of the current sequence points and is up to 4096).
M Sequence: a pseudo-random sequence with a 4096 length.
m Sequence: a pseudo-random sequence with a 4096 length.
2.
Stored Sequence
Press Load Sequence
Stored to open Store/Recall interface, select and read
the stored FH sequence file from the internal or external memory of the
instrument. For details, please refer to the introduction in “Store and Recall”.
3.
Volatile Sequence
Press Load Sequence
Volatile to select the sequence from the current
volatile memory.
n
s
1
s
3
s
8
s
n
www.
.com
1.800.561.8187
Summary of Contents for DG5071
Page 2: ...www com information itm com 1 800 561 8187...
Page 46: ...www com information itm com 1 800 561 8187...
Page 78: ...www com information itm com 1 800 561 8187...
Page 114: ...www com information itm com 1 800 561 8187...
Page 126: ...www com information itm com 1 800 561 8187...
Page 134: ...www com information itm com 1 800 561 8187...
Page 200: ...www com information itm com 1 800 561 8187...
Page 210: ...www com information itm com 1 800 561 8187...