Chapter 10 Advanced Operation
RIGOL
User’s Guide for DG5000
10-5
Sync Polarity
Set the sync signal on the [Sync] connector as normal or invert.
Press Utility
CH1 Set
Polarity to select “Positive” or “Negative”. Notice that
the sync signal related to the waveform is not invert even when the waveform is
invert.
Output Polarity
Set the signal on the [Output] connector as normal or invert. Waveform inverts
relatively to the offset voltage.
Press Utility
CH1 Set
Output to select “Normal” or “Invert”. The default is
“Normal”. The first half period of the cycle waveform is positive in “Normal” mode and
negative in “Invert” mode.
0 V
Offset
0 V
Normal Invert
Normal Invert
After the waveform is inverted, please pay attention to the following cases
:
None offset voltage is changed.
The waveform viewed in Graphic mode is not invert.
The sync signal related to the waveform is not invert.