Chapter 3 Arbitrary Waveform Output
RIGOL
User’s Guide for DG5000
3-17
Edit Block
The following example introduces how to create an arbitrary waveform which meets
the following conditions with the following conditions met by using “Edit Block”.
Parameter
Value
Cycle Period
12 μs
High VLimit
4 V
Low VLimit
-2 V
Interpolation
Linear
Start
2, 4 V
End
4, -2 V
Operating Procedures:
1.
Press Arb, use
to open the 2/2 menu page, and then press Create New
Cycle Period, use the numeric keyboard to input “12” and select the unit “μs”
in the pop-up menu.
2.
Press High VLimit, use the numeric keyboard to input “4” and select the unit “V”
in the pop-up menu.
3.
Press Low VLimit, use the numeric keyboard to input “-2” and select the unit “V”
in the pop-up menu.
4.
Press Init #Points, use the numeric keyboard to input “4” and then press OK.
Note that a level line of -2 V appears.
5.
Press Interp to select “Linear”.
6.
Press Edit Block to enter the “Edit Block” interface.
1)
Press X1 and use the numeric keyboard to input “2”. Then press Y1 and use
the numeric keyboard to input “4” and select the unit “V” in the pop-up menu.
Use the same method to set X2 and Y2.
2)
Press Excute to apply the setting defined in step 1). The waveform editor will
www.
.com
1.800.561.8187
Summary of Contents for DG5071
Page 2: ...www com information itm com 1 800 561 8187...
Page 46: ...www com information itm com 1 800 561 8187...
Page 78: ...www com information itm com 1 800 561 8187...
Page 114: ...www com information itm com 1 800 561 8187...
Page 126: ...www com information itm com 1 800 561 8187...
Page 134: ...www com information itm com 1 800 561 8187...
Page 200: ...www com information itm com 1 800 561 8187...
Page 210: ...www com information itm com 1 800 561 8187...