RIGOL
Chapter 10 Advanced Operation
User’s Guide for DG5000
10-42
Press ManualIP to select “On” or “Off” to turn ManualIP mode on or off. To
enable this mode, set DHCP and AutoIP as “Off”.
The IP format is nnn.nnn.nnn.nnn. The first nnn ranges from 0 to 223
(except 127) and the other three nnn range from 0 to 255. We recommend
that users acquire an available IP from your network administrator.
Press IP Address and use the numeric keyboard and direction keys to enter
your desired IP. The setting is stored in the non-volatile memory and will be
loaded automatically when the generator is powered on the next time if
DHCP and AutoIP are set as “Off”.
Tips
If all the three IP configure modes are “On”, the priority of parameters
configuration from high to low is “DHCP”, “AutoIP” and “ManualIP”.
The three IP configure modes cannot be set as “Off” at the same time.