RBP-
100
4.2.3 There is an interval between measurements. A count down indicates the remaining time,
then the device repeats the measurement. If beeper is activated beep will sound when
count down reaches 5 seconds.
4.2.4 After the result of the second measurement is displayed, the device performs again a
countdown and again repeats the measurement.
4.2.5 When the whole triple BP measurement episode is completed, the average is calculated.
The display then shows the average result, comprising the systolic (17) and the diastolic
(18) blood pressure, pulse rate (19) and mean arterial pressure (MAP) (22). The results of
pulse rate and mean arterial pressure alternating appear on the display every 2 seconds.
Note:
- To skip an interval count down, press the Start/Stop button (3) during count down.
Measurement will be initiated immediately.
- The bottom hand section of the display shows “N= “1, 2 or 3 to indicate which of the 3
measurements is currently being taken.
- If one of the individual measurements was questionable, a fourth one is automatically taken.
- Do not remove the cuff between measurements.
- The length of the interval time between the measurements can be adjusted (see chapter 2.5)
- Manual inflation: If the systolic blood pressure of a patient is known to be very high,
it is possible to set the pressure individually in each measurement. Press the Plus button
(5) after the monitor has been pumped up to a level of approx. 30 mmHg (shown on the
display). Keep the button pressed until the pressure is about 40 mmHg above the expected
systolic value – then release the button.
- Manual rapid cuff deflation: Press and hold the Minus button (6) when you want to deflate
the cuff rapidly.
- You can stop the measurement at any time by pressing the ON/OFF button (e.g. if your
patient feel uneasy or an unpleasant pressure sensation).
4.3 AUSCULTATORY/MANUAL BLOOD PRESSURE
MEASUREMENT (MAN)
4.3.1 Select the manual BP measurement mode by pressing repeatedly the Mode button (2)
until MAN (32) appears on the display.
4.3.2 Briefly press the Start/Stop button (3) to start the automatic inflation of the cuff (28). The
cuff will be inflated automatically to 30-40mmHg above systole. Alternatively, press and
hold “+” (5) button after cuff pressure reaches 40mmHg to manually inflate the cuff to
pressure of your choice; release “+” (5) button to stop manual inflation.
4.3.3 Once peak pressure is reached, the cuff will deflate at 3mmHg per second. The deflation
is indicated by the cuff deflation symbol (27).
Summary of Contents for 04045396208797
Page 1: ...RBP 100 Bedienungsanleitung DIGITALES BLUTDRUCKMESSGERÄT 0124 ...
Page 31: ...RBP 100 Instruction for use DIGITAL BLOOD PRESSURE MONITOR 0124 ...
Page 61: ...RBP 100 Instructions d utilisation TENSIOMÈTRE NUMÉRIQUE 0124 ...
Page 91: ...Instrucciones de uso MONITOR DE PRESIÓN ARTERIAL DIGITAL 0124 RBP 100 ...
Page 121: ...RBP 100 Инструкция по эксплуатации ЦИФРОВОЙ ТОНОМЕТР 0124 ...
Page 151: ...RBP 100 Istruzioni per l uso MISURATORE DIGITALE DELLA PRESSIONE SANGUIGNA 0124 ...
Page 181: ...RBP 100 ...
Page 182: ...RBP 100 ...
Page 183: ...RBP 100 ...