RBP-
100
3.1.1 Always ensure that the correct cuff size is used (size markings on the cuff)
3.1.2 Contact your local Riester service if the enclosed cuffs (12) do not fit.
3.1.3 Connect the cuff to the device by clicking the cuff connector (13) on to the cuff socket (8).
Only use Riester cuffs.
3.2 FITTING THE CUFF
3.2.1 Remove close-fitting garments from the upper arm of the patient. To avoid constriction,
shirt sleeves should not be rolled up - they do not interfere with the cuff if they are laid
flat.
3.2.2 Position the cuff on the upper arm (right or left), so that the tube points in the direction
of the lower arm.
3.2.3 The artery mark on the cuff must lie over the artery which runs down the inner side of
the arm.
3.2.4 Make sure that the cuff is positioned 2-3 cm above the elbow.
3.2.5 Secure the cuff with the Velcro and make sure the cuff is comfortably attached and not
too tight.
3.2.6 Lay the arm of the patient on the table (palm upwards), so that the cuff is at the same
height as the heart.
3.2.7 Make sure that the tube is not kinked.
3.3 SELECTING THE MEASUREMENT MODE
This device allows you to measure in three different measurement modes. Select between
Standard BP Measurement 1x (30), Triple BP Measurement 3x (31) and Manual BP Measurement
(32) by pressing repeatedly the Mode button (2). The current set measurement mode is on the
display shown by the related symbol.
Seite 10 von 23
•
The device is delivered with default setting activated beeper.
3.
Before each measurement
3.1 Selecting the correct cuff
Riester offers different cuff sizes. Select the cuff size to match the circumference of your patient's
upper arm (measured by close fitting in the centre of the upper arm).
3.1.1 Always ensure that the correct cuff size is used (size markings on the cuff)
3.1.2 Contact your local Riester service if the enclosed cuffs (12) do not fit.
3.1.3 Connect the cuff to the device by clicking the cuff connector (13) on to the cuff socket (8).
Only use Riester cuffs.
3.2 Fitting the cuff
3.2.1 Remove close-fitting garments from the upper arm of the patient. To avoid constriction, shirt
sleeves should not be rolled up - they do not interfere with the cuff if they are laid flat.
3.2.2 Position the cuff on the upper arm (right or left), so that the tube points in the direction of the
lower arm.
3.2.3 The artery mark on the cuff must lie over the artery which runs down the inner side of the arm.
3.2.4 Make sure that the cuff is positioned 2-3 cm above the elbow.
3.2.5 Secure the cuff with the Velcro and make sure the cuff is comfortably attached and not too
tight.
3.2.6 Lay the arm of the patient on the table (palm upwards), so that the cuff is at the same height
as the heart.
3.2.7 Make sure that the tube is not kinked.
3.3 Selecting the measurement mode
This device allows you to measure in three different measurement modes. Select between Standard
BP Measurement 1x (30), Triple BP Measurement 3x (31) and Manual BP Measurement (32) by
pressing repeatedly the Mode button (2). The current set measurement mode is on the display shown
by the related symbol.
4.
Performing BP measurements in different measurement modes
4.1 Standard blood pressure measurement (1x)
4.1.1 Select the standard BP measurement mode by pressing repeatedly the Mode button (2) until
1x (30) appears on the display.
4.1.2 Press the Start/Stop button (3) to start the measurement.
4.1.3 The cuff will now pump up automatically. The inflation is indicated by the blinking cuff inflation
symbol (28). The patient should relax, should not move and should not tense his arm muscles
until the measurement result is displayed. He should breathe normally and not talk.
4.1.4 When the correct pressure is reached, the pumping stops and the pressure falls gradually.
The deflation is indicated by the blinking cuff deflation symbol (27). If the required pressure
was not reached, the device will automatically pump some more air into the cuff.
4.1.5 During the measurement, the pulse indicator (25) flashes in the display and a beep sound
appears each time a heartbeat is detected.
4.1.6 The result, comprising the systolic (17) and the diastolic (18) blood pressure, pulse rate (19)
and mean arterial pressure (MAP) (22). The results of pulse rate and mean arterial pressure
alternating appear on the display every 2 seconds.
4.1.7 When the device has finished measuring, remove the cuff.
4.1.8 Switch off the device (The monitor does switch off automatically after approx. 1 min.).
Cuff size
For circumference of upper
arm
S (option)
14-22 cm (5.5 – 8.7 inches)
M
22 - 32 cm (8.7 - 12.6
inches)
L-XL
32 – 52 cm (12.6 - 20.5
inches)
Summary of Contents for 04045396208797
Page 1: ...RBP 100 Bedienungsanleitung DIGITALES BLUTDRUCKMESSGERÄT 0124 ...
Page 31: ...RBP 100 Instruction for use DIGITAL BLOOD PRESSURE MONITOR 0124 ...
Page 61: ...RBP 100 Instructions d utilisation TENSIOMÈTRE NUMÉRIQUE 0124 ...
Page 91: ...Instrucciones de uso MONITOR DE PRESIÓN ARTERIAL DIGITAL 0124 RBP 100 ...
Page 121: ...RBP 100 Инструкция по эксплуатации ЦИФРОВОЙ ТОНОМЕТР 0124 ...
Page 151: ...RBP 100 Istruzioni per l uso MISURATORE DIGITALE DELLA PRESSIONE SANGUIGNA 0124 ...
Page 181: ...RBP 100 ...
Page 182: ...RBP 100 ...
Page 183: ...RBP 100 ...