24
VPAP® III, VPAP III ST, VPAP III ST-A Service Manual
11 Bend tube 3 to the front of the unit, as shown below:
12 Turn the blower assembly over and install it in the lower case, taking care with the location of
the blower inlet connector and blower outlet connector. Pay special attention to the routing of
the silicon tubes so that no kinking or compression occurs.
Note:
Ensure the three vibration mounts are sitting firmly between the blower module and
the lower case.
Blower inlet connector
slides onto inlet seal post
Blower outlet connector
slides onto post
Location of the vibration
mounts