7 Calibration procedures
59
3333
Motor noise calibration (VPAP III ST-A only)
Motor noise calibration (VPAP III ST-A only)
Motor noise calibration (VPAP III ST-A only)
Motor noise calibration (VPAP III ST-A only)
Caution
The flow calibration must be performed before the motor noise calibration.
1 Select
Calibration>Motor noise calibration
,
or press the
Motor noise calibration
button at the top of the main window.
2 Follow the prompts on screen. Once the calibration is complete the following screen will
appear:
Once the calibrations are complete, repeat tests 6 - 12 from
Section 3 General inspection
and servicing
.
If the unit fails calibration, ensure the test set-ups are correct. If the unit continues to fail, send the
flow generator with product problem report (found on page 65 of this manual) to ResMed for
further servicing.
!