58
VPAP® III, VPAP III ST, VPAP III ST-A Service Manual
4 Follow the prompts on the screen. Once the calibration is complete the following screen will
appear:
2222
Flow calibration
Flow calibration
Flow calibration
Flow calibration
Caution
The pressure calibration must be performed before the flow calibration.
1 Select
Calibration>Flow calibration
,
or press the
Flow calibration
button at the top of
the main window.
2 Follow the prompts on the screen. Once the calibration is complete the following screen will
be displayed:
!