5 Replacing major subassemblies
41
2 Take the AC switch label (R241-406) and place it on the outside of the bottom case, next to
the AC switch, as shown below:
Note:
The AC switch label is not required for Teijin VPAP III ST-A units.
Replacing interface assembly components
1 Turn the interface assembly upside down and release the clips holding in the LCD module.
Take care not to touch the gold contacts on the back of the LCD module.
AC switch label
Clips
Clips
Do not touch the gold contacts