20
VPAP® III, VPAP III ST, VPAP III ST-A Service Manual
5 Turn the module upside down and release the locking tabs that are holding the flexible cable.
6 Gently pull out the flexible cable.
Remove the filter
7 Take hold of the lower end of the air filter cover and pull it out and away from the unit.
Remove the top case
8 With the interface assembly removed, turn the unit upside down and undo the four screws
holding on the top case, using a Phillips head screwdriver.
Locking tabs
Screws
Screws