3 General inspection and servicing
15
14
14
14
14 Reload clinical settings
Reload clinical settings
Reload clinical settings
Reload clinical settings
1 Select
Settings>Flow generator settings
from the main window.
2 Select
File>Read settings from file
, or select the folder icon.
3 Navigate to where the settings file was saved and open the file.
4 Once the settings are displayed on the screen, click the
Set
button to upload them to the flow
generator.
5 Close the Flow generator settings window.
15
15
15
15 Complete the Service record results form
Complete the Service record results form
Complete the Service record results form
Complete the Service record results form
1 Press the
Service record results form
button and open the file with the unit’s settings.
2 Press the Service details tab.
• Check the Reload clinical settings box at the base of the window.
• Check the Clean flow generator box to indicate if the unit was cleaned according to the
procedures outlined in step
1 Clean the flow generator
.
• Check the appropriate HumidAire 2i test box.
• Check the appropriate DC power test box.
• Check the appropriate VPAP III ST-A alarm test box.
16
16
16
16 Fill out product problem report
Fill out product problem report
Fill out product problem report
Fill out product problem report
1 If any problems are found with the flow generator, fill out the
Service request form
in Appendix A
of this manual, and send it to ResMed in Sydney for review and logging. (The address may be
found on the first page of this manual.)
2 Check the appropriate Fill out product problem report box in the Service details window.
3 Save the file. If a hard copy of the form is needed, press the
button at the top of the
window.
4 Close the Record of Service window.