VPAP® III, VPAP III ST, VPAP III ST-A Service Manual
2 Remove the LCD module keypad, taking care not to touch the pills on the keypad.
3 Insert the keypad into the interface panel, once again taking care not to touch the pills on the
keypad.
4 Fit the LCD module to the interface panel by engaging the clips on the right and left of the
interface panel. If this is a new LCD module, remove the protective backing from the LCD
module window, taking care not to touch the gold contacts.
Closing the flow generator
1 Take the flexible connector and carefully feed it through the top case.
2 Turn the unit upside down and fix the top case to the bottom case with the Phillips head
screws. Screw in the rear screws first.
Caution
Thread cutting screws are used in the flow generator. When replacing a thread cutting
screw, turn the screw anti-clockwise until it fits into the previously cut thread, otherwise the
original thread may be stripped out.
Do not touch the pills
(VPAP III keypad pictured)
!