62
VPAP® III, VPAP III ST, VPAP III ST-A Service Manual
9
System specifications
VPAP III, VPAP III ST
Dynamic pressure characteristics
IPAP: 2 cm H
2
O to 25 cm H
2
O (measured at the end of standard 2 m air tubing)
EPAP: 2 cm H
2
O to 25 cm H
2
O (measured at the end of standard 2 m air tubing)
Maximum single fault pressure:
40 cm H
2
O
Dynamic flow characteristics
130 L/min at 2 to 20 cm H
2
O
Sound pressure level:
< 30 dB (tested in accordance with the requirements of ISO
17510-1:2002)
Dimensions (L x W x H):
270 mm x 230 mm x 141 mm
Weight:
2.3 kg
Air outlet:
22 mm taper, compatible with EN 1281-1:1977 Anaesthetic & Respiratory Equipment
- Conical Connectors
Pressure measurement:
Internally mounted pressure transducer
Flow measurement:
Internally mounted flow transducer
Power supply:
Input range 100-240V, 50-60 Hz, 40VA (typical power consumption), <100VA
(maximum power consumption)
Housing construction:
Flame retardant engineering thermoplastic
Environmental conditions
Operating temperature: +5°C (41°F) to +40°C (104°F)
Operating humidity: 10%-95% non-condensing
Storage and transport temperature: -20°C (-4°F) to +60°C (140°F)
Storage and transport humidity: 10%-95% non-condensing
Electromagnetic compatibility
Product complies with all applicable electromagnetic compatibility requirements (EMC) according
to IEC60601-1-2, for residential, commercial and light industry environments.
Air filter:
Two layered, powder-bonded, polyester non-woven fiber
Air tubing:
Flexible plastic, 2 m or 3 m length
IEC 60601-1 Classifications
Class II (double insulated), Type CF
Note:
The manufacturer reserves the right to change these specifications without notice.
Symbols which may appear on the product
i
Start/S
top
Mask-fi
t
Attention, consult accompanying documents
Class II equipment
Type CF equipment
Drip proof
Start/Stop or Mask-Fit