47
WR14 / WR50-12/13 Winding Resistance Meter Instruction Manual
Version 3.03
Quality Assistant:
If the result is within the selected parameters, a green light in the main screen will indicate this condition
and automatically store the result. In the interval mode, the selected Interval measurement will start
recording results.
If the result is out of the selected parameters, a red light in the main screen will indicate this condition.
No results are stored.
If the selected conditions are almost met, a yellow light in the main screen will illuminate and indicate
this condition. No results are stored.
Auto Print Results
8
If this feature is selected, the instrument will also output to a connected printer when the measurements
are stored.
AutoSave USB:
The results will also be stored on a USB Key, if connected to the instrument during a measurement.
8
Not available in the WR14.