OPERATION ELEMENTS FOR FIELD UNITS
WR50
4
FRONT PANEL OVERVIEW:
WR50 CONTROL ELEMENTS AND CONNECTIONS
FUSE, MAINS INPUT AND SWITCH
A 16A fuse (slow blow) protects the device. Use only the correct fuse type to replace it.
EMERGENCY STOP
Press to Stop. The measurement will stop and immediately begin to discharge the transformer. Turn
clockwise to release. The device will remain in a safe mode state.
THERMAL PRINTER
Allows the user to quickly get results on paper by using the built-in thermal printer. The printer uses
thermal heat transfer technology. Therefore, there are no ink cartridges required or printer heads.
Special thermal paper is required for the printer to operate properly.
4
This illustrates a WR50-12 (2 channel instrument), the WR50-13 has a 3
rd
channel.
1: Fuse, Mains Input
and Switch
9: Ground Terminal
2: Emergency Stop
6: Potential
Receptacles
7: Current
Receptacle
10: Temperature
Receptacles 1 -3
8: External
11: Interface
(USB and RS 232)
3: Thermal Printer
4: Safety Indicators
5: Display with
Touch Panel