WR14 / WR50-12/13 Winding Resistance Meter Instruction Manual
22
GROUND TERMINAL
Use this terminal to connect an additional ground line even if your mains Power cable (Mains) provides
an Earth ground line. Also, use it even if you are not sure that the local installation Earth ground is a low
resistive ground.
EXTERNAL
This port is used for an external warning device. See the chapter on interfaces for more details.
TEMPERATURE RECEPTACLE
These two (2) receptacles are for the external temperature probes (Optional accessory part number:
2021N-26001).
POTENTIAL RECEPTACLE
Connect positive (+) and negative (-) potential cables for each channel to the test object.
CURRENT RECEPTACLE
Connect positive (+) and negative (-) current cables for each channel to the test object.
SAFETY INDICATOR
Green Light
on:
System is in safe standby mode. Ready for handling test leads and connections.
Red Light
on:
Unsafe mode! Do NOT remove any cables. There is current flowing and voltage on
the test leads. System is testing.
WARNING!
The transformer is not discharged!
EMERGENCY STOP
Press to Stop. The measurement will stop and immediately begin to discharge the transformer. Turn
clockwise to release. The device will remain in a safe mode state.
MAINS INPUT
Connect a power cord to the Mains Input (88 to 264 V / 47 to 63 Hz).