Version 3.03
WR14 / WR50-12/13 Winding Resistance Meter Instruction Manual
114
TROUBLE SHOOTING
When all WR series instruments are powered on, an internal calibration and check sequence is
performed.
Upon completion of the check sequence, the test set wi
ll proceed to the “Main Screen.”
Should there be any problem detected with the test set, an error message will appear.
System does not display anything:
1. Check the display for any Initialization.
2. Check the fuses. The fuses are located on the rear or the front panel of the instruments.
Touch Panel does not work:
Connect a USB mouse to the USB port and control the instrument with it. The instrument will
automatically display a cursor and everything can be done by mouse.
See “Mouse Cursor” on page 34.
Try to recalibrate the tou
ch panel: See “Calibrate Touchscreen” on page 97.
USB Memory Stick does not work:
There are a few unsupported memory sticks available on the market. Please use another model or
brand and try again.
Test Current cannot be turned on:
Is the Emergency Stop switch pushed in? Turn it clockwise to release.