INSTALLATION MANUAL
–
RAVAS-
SafeLoad
30
7.3. Moment parameters
o
The display will show [P_00
0
] with the most right digit blinking.
Press
the
↑
-key shortly to change the parameter selection.
o
The display will show [P_00
1
] with the most right digit blinking.
Select the proper parameter by using the
↑
-key, the
↓
-key and the
-key.
Moment parameters
Parameter Function
Default
Settings
1
Calibration and start-up
units
1
1=kg;2=lb;3=t;4=%
8
zero track filtering band
10
0;5;10;20,50 % of full span/second
11 display
service
mode
basic basic;count;res10
12
Power On Zero
no
yes; no
14
Fork segments active
ON
OFF/On
15
units switch active
yes
no;yes
16
Zero track timer moment
sensor
4
0 to 9 seconds
19 Absolute
zero
range
positive
500
0 to 500 % of span(reference point
calibrated zero)
20 Absolute
zero
range
negative
500
0 to 500 % of span(reference point
calibrated zero)
49
underload % of FS
100
0-100% of full span
52
Switch point 1
st
to 2
nd
fork
segment
30
0-100 % of range
53
Switch point 2
nd
to 3
rd
fork
segment
60
0-100% of range
54
Correction factor 1
st
segment line
000 0-100%
compensation.
55
Correction factor 3
rd
segment line
000 0-100%
compensation
65
auto shut off backlight
0
0;20; 40; 80; 160; 320 seconds
66 backlight
brightness
150 200/150/100/75/50/25/0
71
Software versie (installed)
x.x
Fixed value
72
Audit trail number
parameters
Changes whenever a parameter is being
changed
73
Audit trail number
calibration
Changes whenever a new calibration is
performed
90
Recover factory default
parameter settings only,
based on actual P01 setting
91
Recover initial parameter
se calibration from
backup (P94)
92
Recover last (re-)calibration
only. And convert calibration
to the actual set startup unit
93
Read out last 50 error
messages(with timestamp)
00-00
Display will show [WAIT] until all errors are
completely loaded.
94
Backup initial parameter
se calibration
00000
95
Measurement on/off
on
On; OFF
97
Key test function (Buzzer
and nr)
99 Software
version
(last
available version)
0.31