![R&S RTA4000 User Manual Download Page 733](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987733.webp)
Remote control commands
R&S
®
RTA4000
733
User Manual 1335.7898.02 ─ 08
POWer:MODulation:RESult:MEAN:PPEak?
POWer:MODulation:RESult:MEAN:STDDev?
POWer:MODulation:RESult:MEAN:WFMCount?
Returns the corresponding statistic result for "Mean".
●
[:ACTual]:
current measurement result
●
AVG:
average of the long-term measurement results
●
NPEak:
negative peak value of the long-term measurement results
●
PPEak:
positive peak value of the long-term measurement results
●
STDDev:
standard deviation of the long-term measurement results
●
WFMCount:
the number of waveforms used for the displayed results
Return values:
<WaveformCount>
Usage:
Query only
POWer:MODulation:RESult:RMS[:ACTual]?
POWer:MODulation:RESult:RMS:AVG?
POWer:MODulation:RESult:RMS:NPEak?
POWer:MODulation:RESult:RMS:PPEak?
POWer:MODulation:RESult:RMS:STDDev?
POWer:MODulation:RESult:RMS:WFMCount?
Returns the corresponding statistic result for RMS.
●
[:ACTual]:
current measurement result
●
AVG:
average of the long-term measurement results
●
NPEak:
negative peak value of the long-term measurement results
●
PPEak:
positive peak value of the long-term measurement results
●
STDDev:
standard deviation of the long-term measurement results
●
WFMCount:
the number of waveforms used for the displayed results
Return values:
<WaveformCount>
Usage:
Query only
POWer:MODulation:RESult:STDDev[:ACTual]?
POWer:MODulation:RESult:STDDev:AVG?
POWer:MODulation:RESult:STDDev:NPEak?
POWer:MODulation:RESult:STDDev:PPEak?
POWer:MODulation:RESult:STDDev:STDDev?
POWer:MODulation:RESult:STDDev:WFMCount?
Returns the corresponding statistic result for the standard deviation.
●
[:ACTual]:
current measurement result
●
AVG:
average of the long-term measurement results
Power analysis (option R&S