![R&S RTA4000 User Manual Download Page 458](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987458.webp)
Remote control commands
R&S
®
RTA4000
458
User Manual 1335.7898.02 ─ 08
0
The number of acquired waveforms is less than the number
required for average calculation. See
1
The instrument acquired a sufficient number of waveforms to
determine the average.
Usage:
Query only
TIMebase:ROLL:AUTomatic
<AutomaticRoll>
Enables the automatic roll mode. The instrument switches to roll mode if the timebase
is equal or slower than the roll mode limit defined with
.
Parameters:
<AutomaticRoll>
ON | OFF
*RST:
OFF
TIMebase:ROLL:MTIMe
<MinTimeBase>
The roll mode is enabled automatically if the time base exceeds the
MinTimeBase
,
and if
Parameters:
<MinTimeBase>
Limit value for roll mode enabling.
*RST:
500e-3
Default unit: s/div
ACQuire:INTerpolate
<InterpolationType>
Defines the interpolation mode.
See also:
Parameters:
<InterpolationType> SINX | LINear | SMHD
LINear
Linear interpolation between two adjacent sample points.
SINX
Interpolation by means of a sin(x)/x curve.
SMHD
Sample & hold causes a histogram-like interpolation.
*RST:
SINX
Waveform setup