![R&S RTA4000 User Manual Download Page 602](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987602.webp)
Remote control commands
R&S
®
RTA4000
602
User Manual 1335.7898.02 ─ 08
SYSTem:COMMunicate:INTerface:USB:CLASs
Selects the USB mode.
●
USB TMC (Test & Measurement Class)
●
USB VCP (Virtual Com Port)
●
USB MTP (Media Transfer Protocol)
Parameters:
<USBClass>
TMC | VCP | MTP
17.10.5
Trigger out
TRIGger:OUT:MODE
<OutputMode>
Defines which signals are generated at the [Aux
Out] connector.
Parameters:
<OutputMode>
OFF | TRIGger | REFerence | MASK
OFF
No output
TRIGger
Outputs a pulse when the instrument triggers.
REFerence
Outputs a 10 MHz reference frequency.
MASK
Outputs a pulse when a mask is violated. This function is only
available if a mask is specified.
*RST:
OFF
TRIGger:OUT:PLENgth
<PulseLength>
Defines the pulse width of the pulse at the [Aux
Out] front connector (at trigger event or
mask viaolation).
Parameters:
<PulseLength>
*RST:
1E-6
TRIGger:OUT:POLarity
<Polarity>
Defines the polarity of the pulse at the [Aux
Out] front connector (at trigger event or
mask viaolation).
General instrument setup