Remote control commands
R&S
®
RTA4000
500
User Manual 1335.7898.02 ─ 08
SEARch:TRIGger:WINDow:RANGe
<RangeMode>
Selects how the signal run is compared with the window.
Parameters:
<RangeMode>
ENTer | EXIT | WITHin | OUTSide | PASSthrough |
NPASsthrough
ENTer
Triggers when the signal crosses the upper or lower level and
thus enters the window made up of these two levels.
EXIT
Triggers when the signal leaves the window.
WITHin
Triggers if the signal stays between the upper and lower level for
a specified time. The time is defined in various ways by the
SEARch:TRIGger:WINDow:TIMerange
OUTSide
Triggers if the signal stays above the upper level or below the
lower level for a specified time. The time is also defined by the
SEARch:TRIGger:WINDow:TIMerange
*RST:
ENT
SEARch:TRIGger:WINDow:TIMerange
<TimeRange>
Selects how the time limit of the window is defined. Time conditioning is available for
the vertical conditions
WITHin
and
OUTSide
).
Parameters:
<TimeRange>
WITHin | OUTSide | SHORter | LONGer
WITHin
Triggers if the signal stays inside or outside the vertical window
limits at least for the time
Width - Delta
and for
Width + Delta
at
the most.
OUTSide
"Outside" is the opposite definition of "Within". The instrument
triggers if the signal stays inside or outside the vertical window
limits for a time shorter than
Width - Delta
or longer than
Width +
Delta
.
SHORter
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
LONGer
Triggers if the signal crosses vertical limits before the specified
"Width" time is reached.
*RST:
LONG
Waveform analysis