Remote control commands
R&S
®
RTA4000
473
User Manual 1335.7898.02 ─ 08
Suffix:
<n>
.
1..5
Parameters:
<Level>
TRIGger:A:LEVel<n>:RISetime:UPPer
<Level>
Sets the upper voltage threshold. When the signal crosses this level, the rise/fall time
measurement starts or stops depending on the selected slope.
Suffix:
<n>
.
1..5
Parameters:
<Level>
TRIGger:A:RISetime:DELTa
<Variation>
Sets a range Δt to the reference rise/fall time set with
,
if
Within
or
Outside
. The instrument trig-
gers on rise/fall times inside or outside the range
time ± Δt
.
Parameters:
<Variation>
TRIGger:A:RISetime:RANGe
<Range>
Sets how the measured rise or fall time is compared with the given limit(s).
To set the rise/fall time, use
.
.
Parameters:
<Range>
LONGer | SHORter | WITHin | OUTSide
LONGer
Triggers on rise/fall times longer than the given time.
SHORter
Triggers on rise/fall times shorter than the given time.
WITHin
Triggers on rise/fall times inside the range
time ± Δt
.
OUTSide
Triggers on rise/fall times outside the range
time ± Δt
.
*RST:
LONG
TRIGger:A:RISetime:SLOPe
<Polarity>
Sets the edge of the slope to be triggered on.
Trigger