Remote control commands
R&S
®
RTA4000
450
User Manual 1335.7898.02 ─ 08
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ZHDRange>
AUTO | MHIGh | MLOW
AUTO
The voltage range is set only at the oscilloscope by adjusting the
vertical scale.
MHIGh
Sets the higher voltage range of the connected probe.
MLOW
Sets the lower voltage range of the connected probe.
*RST:
AUTO
PROBe<m>:SETup:ACCoupling
<ProbeCouplingAC>
Enables AC coupling in R&S
RT-ZPR probes, which removes DC and very low-fre-
quency components. The R&S
RT-ZPR probes requires 50
Ω
input termination, for
which the channel AC coupling is not available. The probe setting allows AC coupling
also at 50
Ω
inputs.
Suffix:
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ProbeCouplingAC> ON | OFF
*RST:
OFF
17.4.6
R&S ProbeMeter
......................................................................................... 450
....................................................................................... 450
....................................................................................... 451
PROBe<m>:PMETer:RESults:SINGle?
.............................................................................451
PROBe<m>:PMETer:RESults:COMMon?
......................................................................... 451
PROBe<m>:PMETer:RESults:DIFFerential?
..................................................................... 451
PROBe<m>:PMETer:RESults:NEGative?
......................................................................... 452
PROBe<m>:PMETer:RESults:POSitive?
.......................................................................... 452
PROBe<m>:PMETer:VISibility
<State>
PROBe<m>:SETup:OFFSwitch
<State>
Switches the integrated voltmeter of a Rohde
&
Schwarz active probe on or off. The
command is only relevant if a Rohde
&
Schwarz active probe with R&S ProbeMeter is
used.
Waveform setup