General instrument setup
R&S
®
RTA4000
235
User Manual 1335.7898.02 ─ 08
Preset does not change the display settings, for example, intensities and brightness.
To reset these settings, restore the factory settings.
To restore the factory settings:
1. Press [Save Load].
2. Tap "Setup".
3. Tap "Factory Default".
If it is not possible to reset the instrument using the keys and the touchscreen, proceed
as follows:
1. Shut down the instrument: press the [Standby] key.
2. Start the instrument: press the [Standby] key.
3. As soon as the message "Press Autoset to restore English language" is shown on
the start screen: Press and hold the Preset key until the "Update firmware" dialog is
shown.
4. Tap "Exit" to close the dialog box.
All instrument settings are reset to factory defaults, and you can use the instrument
as usual.
11.4
Locking the touchscreen
The Touch Lock key locks the touchscreen to prevent unintended use. When the
touchscreen is off, the key is illuminated. Press again to unlock the touchscreen.
11.5
Performing a self-alignment
The self-alignment aligns the data from several input channels vertically and horizon-
tally to synchronize the timebases, amplitudes and positions.
Recommendation on performing the self-alignment:
●
When putting the instrument into operation for the first time
●
After a firmware update
●
Once a week
●
When major temperature changes occur (> 5°)
Performing a self-alignment