![R&S RTA4000 User Manual Download Page 153](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987153.webp)
Measurements
R&S
®
RTA4000
153
User Manual 1335.7898.02 ─ 08
Meas. type
Symbol
Description
Graphic / formula
Mean Cycle
MeanCyc
in V
Mean value of the left-most signal period.
DC
N T
RMS Cycle
RMS-Cyc
in V
RMS (root mean square) value of the voltage of the left-
most signal period.
DC
AC
N T
Peak Peak
Vpp
Difference of maximum and minimum values.
Max
Min
Pk-Pk
Peak+
Vp+
Maximum value within the displayed waveform.
Max
Peak-
Vp-
Minimum value within the displayed waveform.
Min
Pos. Overshoot
Neg. Overshoot
+Ovr
-Ovr
in %
Overshoot of a square wave after a rising or falling edge. It
is calculated from measurement values top level, base level,
local maximum, local minimum, and amplitude.
%
100
Amplitude
Top
Max
Over
local
%
100
Amplitude
Min
Base
Over
local
Top
Base
Amplitude
Over+
Over-
Mean Value
Mean
Arithmetic average of the complete displayed waveform.
N
k
k
x
N
1
1
Mean
RMS Value
RMS
RMS (root mean square) value of the voltage of the com-
plete displayed waveform.
N
k
k
x
N
1
2
1
RMS
σ
-Std. Deviation
σ
Standard deviation of the displayed waveform.
N
k
k
x
N
1
2
Mean
1
1
σ
Automatic measurements