- 70 -
Procedure:
1. Press <
Air density> button.
2. Menu of Air Density function is displayed.
3. Press field for entering weight value of steel mass standard, enter the value as
provided in calibration certificate.
4. Press field for entering weight value of aluminium mass standard, enter the value as
provided in its calibration certificate.
5. Press field for entering density value of steel mass standard, enter the value.
6. Press field for entering density value of aluminium mass standard, enter the density
value.
7. Press START field, air density determination begins.
8. Load the pan with steel mass standard, wait for a stable measurement result and
press <
> button.
9. Load the pan with aluminium mass standard, wait for a stable measurement result
and press <
> button.
10. Density result is displayed.
11. Press <
> button to complete the procedure.
The value of determined air density is automatically assigned to <Air density> parameter in
<Settings/ Air buoyancy compensation> menu for weighing working mode.